Xylem wtw CarboVis 701 IQ TS Operating Manual page 33

Optical sensor for carbon sum parameters and total suspended solids (tss)
Table of Contents

Advertisement

CarboVis 70x IQ TS
Flowchart
ba75950e03
05/2017
Cleaning the sensor
Sensor check
(H2O dist.)
Assessment number
-40 ... +40
< -40 or
> +40
Zero adjustment
Error
OK
Sensor check
(Standard)
OK
Error
Sensor OK
Main sensor:
Check user calibration
Fig. 4-3
Sensortest/Zero adjustment and user calibration.
Overview of
TEST: MEASUREMENT WINDOWS
Clean the measurement windows
Rinse the measuring chamber
with ultrapure water several times
Carry out a sensor check (H2O)
After cleaning several times
without significant change of
the sensor assesssment:
Carry out zero adjustment
Operating error or defective sensor
(Sensor blocked for measurement)
After several zero adjustments:
Reset to:
Factory zero adjustment or
last valid user zero adjustment
TEST: ELECTRONICS AND OPTICS
After several sensor checks:
Operating error or defective sensor
Virtual & secondary sensor:
Check user calibration
Measurement / Operation
33