CarboVis 70x IQ TS
TSS measurement,
5 mm gap width
Suitability and areas
of application
Measuring medium
Pressure resistance
Type of protection
Immersion depth
Operating position
ba75950e03
05/2017
Measuring location
Outlet
The quoted measuring ranges are nominal measuring ranges that
are theoretically possible. In practice, real measuring ranges exist
that are given by the limits of photometric determination. The limits
are significantly influenced by the light scattering due to solids and
the absorption of accompanying substances (sample matrix).
Therefore, no fixed measuring range can be determined before-
hand.
8.2
Application characteristics
Municipal wastewater with a low proportion of industrial wastewater
pH value
Flow rate
Temperature
Sensor with connected SACIQ (SW) sensor connection cable:
Max. admissible overpressure
The sensor meets all requirements according to article 3(3) of 97/23/EC ("pres-
sure equipment directive").
Sensor with connected SACIQ (SW) sensor connection cable:
5
IP X8, 1 bar (10
Pa)
min
max
Any
Measuring mode
TSS
4 - 12
< 3 m/s
0 °C ... + 45 °C
(32 °F ... + 113 °F)
105 Pa (1 bar)
The measuring gap must be
filled with test sample.
10 m depth
Measuring range
0.0 ... 900.0 mg/L
Technical data
Resolution
0.1 mg/L
49