7.10 Automatic adding-up

This function is used to add the individual weighing values automatically into the
summation memory and to provide a printout via the installed printer.
To set this function, see chap. 8.2 „AU on"
 Place material to be weighed A
After stabilisation display has taken place, you will hear an
audio sound.
The symbol for „AUTO" appears.
The weighing value will be saved to the summation memory.
 Remove material to be weighed A
„ACC 1"followed by the weight value will be displayed shortly
and printed.
Afterwards the weighing scales change to zero display.
 Place material to be weighed B
After stabilisation display has taken place, you will hear an
audio sound.
 The weighing value will be saved to the summation memory.
 Remove material to be weighed B
„ACC 2"followed by the total weight of both weighings will be
displayed shortly.
The weight of the items to be weighed B will be printed.
After that the balance changes to the zero display.
 Add more material to be weighed as described before.
Please note that the balance must be unloaded between
the individual weighing procedures.
 You can repeat this process until the capacity of the scales
is exhausted.