Markings
I3 China Intrinsic Safety
Certificate
Standards
Markings
7.9 
Korea
EP Korea Flameproof
Certificate
Markings
Special Conditions for Safe Use (X):
See certificate for special conditions.
IP Korea Intrinsic Safety
Certificate
Markings
Special Conditions for Safe Use (X):
See certificate for special conditions.
7.10  Japan
E4 Japan Flameproof
Certificate
Markings
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm less than 1 mm
thick that forms a boundary between EPL Ga (process
connection) and EPL Gb (all other parts of the equipment). The
model code and datasheet are to be consulted for details of
36
Pressure transmitter: Ex db IIC T6···T4 Ga/Gb
Flow meter: Ex d II C T6~T4 Ga/Gb
GYJ22.1834X; GYJ20.1487X [Flow meters]
GB3/T 3836.1-2021, GB/T 3836.4-2021, GB3836.20-2010
Ex ia IIC T4 Ga, FISCO :Ex iaⅡCT4 Ga,Ex db+ib/ibⅡCT4
Ga/Gb
12-KB4BO-0342X, 12-KB4BO-0344X, 19-
KB4BO-0978X
Ex d IIC T6...T4 Ga/Gb, T4/T5 (-60 °C ≤ Ta ≤ +80 °C),
T6 (-60 °C ≤ Ta ≤ +70 °C)
12-KB4BO-0343X, 12-KB4BO-0345X, 13-
KB4BO-0205X, 13-KB4BO-0207X, 18-KA4BO-0309X
Ex ia IIC T4 (-60 °C ≤ Ta ≤ +70 °C)
CML20JPN112X
Ex db IIC T6...T4 Ga/Gb, T6 (-60 °C ≤ Ta ≤ +70 °C),
T5/T4 (-60 °C ≤ Ta ≤ +80 °C )