CLEANING AND DISINFECTION
B. Braun SpaceStation
All electronic connections and edp links have to be disconnected
from the power grid before cleaning or disinfection of the Space
Station.
Examine the SpaceStation on visual contamination and clean
with a pre-impregnated moist – NOT wet, fleece cloth.
Dispose the cloth after usage.
Our recommendation: ready-to-use Meliseptol
or Meliseptol
Foam pure.
®
In case of heavy contamination remove cover – unlock the lock-
ing button and remove the SpaceCover. Remove visual contami-
nation on the SpaceCover and the guide rail as described above.
Dispose the cloth after usage.
In order to improve the cleaning / disinfection, repeat the cleaning
and disinfection step.
Make sure that no liquid is leaking into the ventilation or con-
nectors. Let the SpaceStation time to dry after cleaning / disin-
fection. Afterwards assemble the SpaceCover again and lock it.
After evaporation of the disinfectant (please observe contact
time) reinsert the clean and disinfected pump into the station
and protect the device from dirt until the next usage.