B. Braun Perfusor Space Service Manual

B. Braun Perfusor Space Service Manual

Hide thumbs Also See for Perfusor Space:

Advertisement

Quick Links

0-12
Perfusor® Space
Service Manual
Version 3.0 English
This document is not a complete Service Manual.
It only encompasses the pages that describe how to
determine the age and how to check the condition
of batteries as well as the accordingly adapted
template for Technical Safety Check.
0 -

Advertisement

Table of Contents
loading

Summary of Contents for B. Braun Perfusor Space

  • Page 1 0-12 Perfusor® Space Service Manual Version 3.0 English This document is not a complete Service Manual. It only encompasses the pages that describe how to determine the age and how to check the condition of batteries as well as the accordingly adapted template for Technical Safety Check.
  • Page 2 Disassembly / Assembly 3.2 Battery Module Designation Ord. No. Battery compartment cover PSP, cpl....3452 0872 Battery module SP (NiMH) without fixing pin......3452 0856 with fixing pin.
  • Page 3 Checklist for Technical Safety Check – Every 24 Months Unit: Perfusor®Space User Manufacturer: B. Braun Melsungen AG Observe the Service Manual and the instructions for use. All measured values are to be documented. Accessories used should be included in testing. Make exclusive use of calibrated measuring equip- ment.
  • Page 4 Technical Safety Check (TSC) Index g (Master – to be added to the documentation) Visual Inspection Electrical Safety Functional Inspection According to IEC / EN 60601-1 or VDE 0750 and VDE 0751 Operation ❒ Syringe fastening ❒ Syringe recognition ❒ Infusion ❒...
  • Page 5 Technical Safety Check (TSC) Index g (Master – to be added to the documentation) Mechanical Aids and Measuring Accessories Used Equipment Used ❒ Syringe gauge, serial No.__________ ❒ Power Supply SP Calibrated until_________________ ❒ Battery module ❒ Service connector SP ❒...
  • Page 6 Procedural Instructions on the TSC Functional Inspection Perfusor® Space Mechanical Inspection 1. Fit the unit to be tested on top of another Space device and check the proper functioning of the lock. 2. Fit the unit to be tested under another Space device and check the proper functioning of the lock.
  • Page 7 Procedural Instructions on the TSC 2. Battery check Use the battery maintenance mode to discharge the battery to battery pre-alarm (the battery status indicator displays no bar). Perform 5 syringe changes with emptied syringe. Replace the battery if a battery alarm occurs during the syringe changes.
  • Page 8 Procedural Instructions on the TSC 7. Trigger bolus by pressing the PCA button. For this purpose, select a PCA drug from the database, fit the PCA button to P2, and trigger bolus. Confirm that the green LED on the PCA button lights up. Bo- lus is delivered.
  • Page 9 Revision Documentation Version 3.0 Changed TSC of the unit Revised procedural instructions on the TSC Release date: 25.10.2013 9 - 2 Perfusor® Space, 3.0...
  • Page 10 B. Braun Melsungen AG | Carl-Braun-Strasse 1 | 34209 Melsungen | Germany Tel +49 (0) 5661 71-0 | Fax +49 (0) 5661 71-4567 | www.bbraun.de Mat. No. M688_001152, Doc. No. 8713 9020, Version 3.0, Effective 25.10.2013...