All electronic connections have to be disconnected from the
power grid before cleaning or disinfection.
Clean / disinfect all sides of the Space Infusion pump.
Start with the display and use ready-to-use wipes or a moistened
fleece cloth. Clean with a moist, NOT a wet fleece cloth.
Dispose cloth after usage.
Our recommendation: ready-to-use Meliseptol
Wipes sensitive
®
or Meliseptol
Foam pure.
®
To clean the flexible cable of the the Infusomat
Space and of
®
the Perfusor
Space we recommend Meliseptol
Foam pure.
®
®
Clean with cotton carrier in use with lint-free tissue.
Make sure that no liquid is leaking into the ventilation or
connectors. Let the device at least one minute to dry after
cleaning / disinfection.
After the drying time reinsert the cleaned and disinfected Space
pump into the station / holder (see cleaning plan of the Space
Station) and protect the device from dirt until the next usage.