Dose To Patient - GE Brivo OEC 715 Operator's Manual

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

1.6.3. Dose to Patient

Skin Dose Levels
The skin dose levels in normal use may be high enough to cause deterministic effects. The availability of the different settings has a considerable
effect on the radiation quality, the delivered dose rate and the image quality.
Patient Entrance Reference Point
Air Kerma (Kinetic Energy Released per unit Mass, i.e. Dose) is an indication of the radiation delivered to the patient entrance reference point, a
point in space, and is measured in the units of Grays (Gy). The System determines the patient entrance reference point as a point along the central
X-ray beam axis 70cm from the focal spot (30cm from the image intensifier entrance).
Selection of the patient entrance reference point is based on international convention and typical usage of a multi-use mobile
NOTE
fluoroscopy system. This point should represent a typical intersection of patient surface and the X-ray beam axis.
Radiation Data
The measurement is performed at patient entrance reference point. Following dosimetric data are typical values in normal use cases. There might
be with some tolerance in actual use cases.
The following data are for Brivo OEC 715, 785 with I.I.(23XZ4ST/BS).
Operation
Technique
Mode
Pediatrics
Mag mode
Normal
kVp
Fluoro
mA
dose rate
(mGy/min)
High Level
kVp
Fluoro
mA
dose rate
(mGy/min)
Auto (PMMA 200mm)
With Grid, No Filter
Mag0
Mag1
Mag2
80
81
80
2.1
3
4
10.33
15.17
19.67
81
81
80
3.2
4.6
6.4
16.18
23.26
31.48
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
With Filter, No Grid
Mag0
Mag1
Mag2
72
73
74
1.7
2.5
3.6
3.36
5.15
7.70
72
73
74
2.5
3.7
5.4
4.94
7.62
11.56
Max
With Grid, No Filter
With Filter, No Grid
Mag0/1/2
Mag0/1/2
110
110
4
4
38.87
24.29
110
110
6
6
58.30
36.43
Page 1-9