Start Up and Operating Control

4.1. Overview

This chapter describes:

Start Up

C-Arm Controls
Workstation Controls
Image Quality Control
4.2. Start Up

4.2.1. Power On

Press the Power On/Off key on the Workstation to start up Workstation, use Workstation to view images.
The C-Arm receives power through an interconnect cable attached to the Workstation, connect C-Arm to
produce images.
At initial setup or when removing from storage, stabilize for 24 hours at ambient
temperature and humidity before applying power. Failure to observe this warning will result
WARNING
in damage to equipment.
Remove any devices from USB connector before booting
fail.
CAUTION
1. Plug the Workstation power plug into a properly rated receptacle. Refer to the Chapter 12 "Technical
Reference" for information about power requirements.
If the system power off automatically due to power lost, user needs to power on the system
after power restored to proceed.
NOTE
Brivo OEC 715/785/865 Mobile C-Arm X-Ray product provides a UPS to protect data loss. If
the power lost during operation, UPS battery will supply power to computer and right
monitor until system shut down automatically.
Do not disconnect, repair or replace UPS without permission. Only a GE qualified service
engineer can perform repair and replacement. The life cycle of the UPS is about 3-5 years in
WARNING
normal use. It may be different in different conditions.
To use the system for the first time, it is recommended to boot the system for continuous 8
hours to ensure that the UPS is fully charged, otherwise if the system power down
accidentally, there is risk that the system will not safely power off; during the normal work,
NOTE
when the system power down accidentally, the UPS play a protective role to ensure the
system safely power off, and it is recommended to boot the system 8 hours in advance to
ensure that the UPS is fully charged the next time; an additional charge of the UPS is not
necessary when the system does not abnormally power off.