4.2.2. Standby or Power Off

Standby:
1. Turn the C-Arm key switch to the standby position (
Placing the equipment in standby disables X-Rays and vertical column operation so these
NOTE
features will not be activated unintentionally.
2. Turn the key switch back to on position when you are ready to use the C-Arm.
Power off:
1. Press power on/off key on Workstation to power off system.
2. Unplug the Workstation power cord from the AC receptacle.
3. Disconnect the Workstation interconnect cable from the C-Arm.
When you turn the system off, wait for at least 1 minute before powering the system back
CAUTION
on. This helps prevent from operational problems.
Remove all power from the system before moving the system or if a problem occurs which
prevents normal operation. All power should also be removed when periodic maintenance
and cleaning is performed.
NOTE
During shutting down, the power shall not be cut off automatically until the system
software exits.
During powering off, the system will ignore power on command by pressing power on/off
NOTE
button until shutdown finished.
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
). The system is still power on.