The Recommended Separation Distances are listed in the next table, which are applicable for Ed3.0 only.
Recommended Separation Distances for Portable and Mobile RF Communications Equipment
and the Brivo OEC 715/785/865 system
The Brivo OEC 715/785/865 Mobile C-arm X-ray Product is intended to use in the electromagnetic
environment where the RF interference is controlled. According to the power rating of the communication
equipment, the purchaser or operator of the Brivo OEC 715/785/865 Mobile C-arm X-ray Product can
prevent the Electromagnetic interference by the recommended separation distance below:
Rated Power of
Transmitter/W
0.01
0.1
1
10
100
For transmitters rated at a power not listed above, the Separation Distance can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE 3: These are guidelines. Actual conditions may vary.
Separation Distance/m
150KHz to 80 MHz
3
5 ,
d
=
[
]
P
3
0.12
0.38
1.2
3.8
12
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
80 MHz to 800 MHz
3
5 ,
d
=
[
]
P
3
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
7
d
=
[
]
P
3
0.23
0.73
2.3
7.3
23