Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
12.5. Scatter Radiation
Isokerma Map Test
This test method is according to clause 203.13.4 in EN 60601-2-43:2010.
The measurement is performed at 2 planes at typical height: 1m & 1.5m from the ground.
The scatter radiation is measured in 2 layouts as vertical and lateral configuration.
For the vertical configuration, the table is at height 1m. I.I. is above the tube.
For the lateral configuration, the table is at height 1m. The center of I.I. is at same altitude as the center of
phantom. The distance from the surface of I.I. to center of phantom is 40cm.
The layout is as below: