GE Brivo OEC 715 Operator's Manual page 185

Mobile c-arm x-ray product
Hide thumbs Also See for Brivo OEC 715:
Table of Contents

Advertisement

Film
Film:
Voltage: 40~110kV;
mAs: 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13, 16, 20, 25, 32, 40, 50, 63, 80 mAs @ 200V, 220V,
230V, 240V;
mAs: 1.0, 1.3, 1.6, 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10, 13, 16, 20, 25, 32, 40 mAs @ 100V, 110V, 120V.
Pulse Fluoro: 1/2/4/8pps
The lowest Current Time Product or the combination of Loading Factors: 40kVp, 1mAs.
Fluoroscopy Duty Cycle:
80 kVp @ 1.0 mA continuous
Film Mode Duty Cycle:
110 kVp @ 20 mA, 4 seconds (80 mAs)
Once per 5 min (1.3%) continuous
Once per 2min (3.3%) for 10 exposures
kVp Accuracy
mA Accuracy
mAs Accuracy
Linearity Film mode linearity
Reproducibility Coefficient of Variation (COV)
Measurement Basis for Technique Factors
kVp - The peak value of high voltage generator output in the interval after a 20 mS delay period to
the end of the exposure.
mA - The time average of the current flow into the high voltage cable/X-ray tube assembly,
beginning at the point where kVp crosses the 35 kVp level.
Time - Measurement of exposure time begins when the kVp crosses the 35 kVp level ( 80% of kVp
selected).
mAs - The time integral of mA as defined above.
Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product
110kV, the obtainable highest X-Ray tube current is 10mA @ 100V, 110V,
120V.
1mAs~80mAs @ 200V, 220V, 230V, 240V; current: 20mA, stationary; the
obtainable highest X-Ray tube voltage is 110kV;
1mAs~40mAs @ 100V, 110V, 120V; current: 10mA, stationary; the
obtainable highest X-Ray tube voltage is 110kV.
7%
10%
(10% + 0.2mAs)
< 0.18
< 0.045
Page 12-7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Brivo oec 785Brivo oec 865

Table of Contents