Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
Table 8: EMC Specifications
The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used in
such an environment.
Emissions test
RF emissions CISPR11
RF emissions CISPR 11
Harmonic Emission
IEC 61000 3-2
Voltage Fluctuations/Flicker
Emission
IEC 61000 3-3
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to EMC information provided in this document.
The Fully Automatic AED Plus is intended for use in the electromagnetic environment specified
below. The customer or the user of the Fully Automatic AED Plus should assure that it is used
in such an environment.
Immunity test
IEC 60601 test
level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV differential
IEC 61000-4-5
mode
+/- 2 kV common
mode
Fully Automatic AED Plus Administrator's Guide
Compliance
Group 1
Class B
Not applicable
Not applicable
Compliance
level
± 6 kV contact
± 8 kV air
Not applicable
± 1 kV I/O
Not applicable
Not applicable
Electromagnetic environment -
guidance
The Fully Automatic AED Plus uses
RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to
cause any interference in nearby
electronic equipment.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.