Appendix B Specifications; General - ZOLL 330 Service Manual

Multifunction aspirator
Hide thumbs Also See for 330:
Table of Contents

Advertisement

General

Operation Modes
Applied Vacuum
Continuous Vacuum Range:
Surgical
Pharyngeal
Tracheal
CASS
Intermittent Vacuum Range ON/OFF Times
FREE AIRFLOW
Collection Canister Capacity
Collection Cannister Material
LED Status/Alarm Indicator
9650-000278-01 Rev. B
Continuous/Intermittent
10 mm Hg to 550 mm Hg
± (2 mm Hg + 8% of setting)
Adjustable 20 to 550 mm Hg
Adjustable 60 to 400 mm Hg
Adjustable 60 to 220 mm Hg
Adjustable 10 to 40 mm Hg
Vacuum Level: Adjustable 10 to 200 mm Hg
ON/OFF Times: 5 to 40 seconds.
Continuous: Up to 30 LPM
Maximum FREE AIRFLOW: Up to 35 LPM*
*Maximum FREE AIRFLOW rate achieved
with the use of Smart Flow.
1200 cc
Polystyrene and durable polyethylene
Red, Yellow, and Green
ZOLL 330 Multifunction Aspirator Service Manual
Appendix B
Specifications
B-1