Instrument Installation; Standard Rack Instrument - Agilent Technologies N5502A Installation Manual

Phase noise measurement system
Table of Contents

Advertisement

6
Preventive Maintenance

Instrument Installation

To install or re- install an instrument in a system, use one of the following
procedures.
Required tools
• #2 Phillips screwdriver
• #2 POZIDRIV screwdriver
• system installation guide

Standard rack instrument

To install an instrument
1 Slide the instrument gently into the rack.
2 Insert the screws in the rack ears.
3 To reconnect the semirigid cables, carefully
4 Turn on system power and restart the system
82
Step
align them before you insert the male
connector.
computer if necessary.
Notes
Most cables are fairly easy to remove and
reconnect and have no special
considerations (besides making sure you
put the cables back in the right place).
Semirigid cables require more care,
especially when reconnecting them. Make
sure all semirigid cables, on the front and
back of an instrument are fully
disconnected before removing the unit.
Do not insert the male pin in at an angle or
you will damage the female connector. RF
connector center pins are very delicate,
and if damaged must be replaced. System
performance may be greatly impaired if
there is a bad RF connector.
E5505A Installation Guide