Agilent Technologies N5502A Installation Manual page 39

Phase noise measurement system
Table of Contents

Advertisement

4
Measurement Software
4
E5500_server_hard_conn
05 Apr 04 rev 1
Figure 15 Select test set, FFT analyzer, and swept analyzer
5
Selecting the three instruments ties those assets to the confidence test performed in the
N O T E
next step.
6
38
In the Asset Manager, verify that the test set and PC Digitizer are
d
configured correctly (check that the license key has been entered
correctly). Also do the following:
Check your system hardware connections.
Click the green check- mark button on the Asset Manager's toolbar to
verify connectivity.
Return to Server Hardware Connections and click the Check I/O
button to re- check it.
From the FFT Analyzer pull- down list, select NI PCI6111- 1 (see
Figure
15). Click the Check I/O button. A green check- mark confirms a
successful I/O check. If the I/O check is not successful, follow the same
process as in
step
3d to verify instrument configuration and connection.
From the Swept Analyzer pull down list select E4411A. Click the
Check I/O button. A green check- mark confirms a successful I/O check.
If the I/O check is not successful, follow the same process as in
to verify instrument configuration and connection.
Close the Server Hardware Connections box.
step
3d
E5505A Installation Guide

Advertisement

Table of Contents
loading

Table of Contents