The sensor should be installed in a location that ensures it remains full during
operation. Vertical installation with upward process fluid flow keeps the
cross-sectional area full, regardless of flow rate. Horizontal installation should be
restricted to low piping sections that are normally full.
Figure 6. Sensor orientation
Mounting position
The electrodes in the sensor are properly orientated when the two measurement
electrodes are in the 3 and 9 o'clock position or within 45 ° from the vertical, as
shown on the right of
measurement electrodes are in the 6 and 12 o'clock position as shown on the left
Figure 7
of
.
Figure 7. Sensor mounting position
FLOW
Figure 7
. Avoid any mounting orientation where the two
Incorrect
FLOW
Quick Start Guide
Correct