Emerson Rosemount 8750WB Quick Start Manual page 21

Magnetic flowmeter system

Advertisement

February 2014
Note
When using the Rosemount supplied combination cable, the electrode wires for terminals 18 and
19 contain an addition shield wire. These two shield wires should be tied with the main shield wire
Figure 17
at terminal 17. See
Figure 17. Combination coil and electrode cable wiring diagram
Coil Drive Cable
1 Red
Integral mount transmitters
The interconnecting wire harness for an integral mount transmitter is installed at
the factory. See
Process Management, Rosemount, Inc.
Figure 18. 8750W integral mount wiring Diagram
.
2 Green 3 Shield
Figure 18
. Do not use cable other than that supplied by Emerson
Quick Start Guide
Electrode Cable
17 Shield 18 Black 19 White
21