CLEANING AND DESINFECTION

8.1

Recommendations

For safety reasons, and to avoid damaging the equipment, the following
recommendations must be observed:
Do not spray liquids directly onto the device, especially the
electrical connectors and the openings of the casing (vents and
air inlet).
Unplug the equipment before cleaning.
Do not use a scraper sponge: risk of damage to the device.
Do not use caustic soda or acetone: the device will be
irreversibly damaged.
Immediately remove any liquid from the equipment with a dry
cloth.
The exterior of the equipment can be cleaned with a sponge or cloth that
has been moistened with water or 70% alcohol.
8.2

Disinfection protocol

Should a tube burst during operation, for example, the parts likely to be
contaminated by infectious agents must be disinfected with a suitable
disinfectant.
In case of possible contamination of the instrument, the external parts
of the instrument (shell, cover, screen, grinding housing) can be cleaned
with a cloth moistened with water and then decontaminated with a
cloth moistened with a bleach solution (1000ppm/6°Cl), or an anionic
disinfectant. The use of ethanol can be considered but should be avoided
on rubber seals.
The user shall assume full responsibility for their choice of disinfection
process.
If another disinfection process must be implemented, contact the
technical support service beforehand in order to confirm that the
procedure is compatible with the equipment.
Page
32
9
MAINTENANCE AND SERVICING
9.1

Repair

The equipment must be cleaned and decontaminated in accordance with
the procedure described in § 8.2 before shipment for repair.
The equipment must be sent with a return form indicating clearly, at the
very least, the procedure used, the persons involved and the date on which
cleaning and/or decontamination was performed. (see appendix 1).
If the return form is missing from the consignment, Bertin Technologies
after-sales team reserves the right to refuse to perform repairs.
9.2

Problems encountered

Encountered issues
Possible cause(s)
There is no power at the mains
socket.
The fan is not turning.
The fuse is defective.
The ventilation system is defective.
There is no power at the mains
socket.
The screen does not turn on.
The fuse is defective.
The display system is defective.
The cap was screwed on incorrectly
or the tube is defective.
The tube used is not a Precellys® tube
A tube is no longer sealed.
The operating limits were not
respected
Power cut
Remedial measures
1.Check the mains voltage.
2.Check that the voltage supplied by
the mains matches that of the device.
3.Check the connection of the device
to the mains.
Replace the fuse.
1.Switch off the device .
2.Contact the technical support..
1.Check the mains voltage.
2.Check that the voltage supplied
by the mains matches that of the
device.
3.Check the connection of the
device to the mains..
Replace the fuse..
1.Switch off the device.
2.Contact the technical support.
If the product contained in the tube
is hazardous or potentially hazardous,
perform the appropriate disinfection
procedure.
For safety reasons, the blocking
plate stays locked in position by the
vacuum system.
Switch the device back on in order to
remove the blocking plate.