Bertin Technologies PRECELLYS Evolution Touch User Manual

Bertin Technologies PRECELLYS Evolution Touch User Manual

Homogenizer

Advertisement

Quick Links

PRECELLYS
®
Evolution Touch
USER MANUAL
PRECELLYS
Evolution Touch
®
WWW.BERTIN-TECHNOLOGIES.COM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PRECELLYS Evolution Touch and is the answer not in the manual?

Questions and answers

Summary of Contents for Bertin Technologies PRECELLYS Evolution Touch

  • Page 1 PRECELLYS ® Evolution Touch USER MANUAL PRECELLYS Evolution Touch ® WWW.BERTIN-TECHNOLOGIES.COM...
  • Page 2: Table Of Contents

    Noise level The information in this document is designed only for use with this Bertin Warranty 28 Technologies product. Bertin Technologies is not liable of any use of this Manufacturer’s Address information that is applied to other products. Technical assistance...
  • Page 3: Technologies

    In the event of non-compliance with the safety instructions set out in The Precellys® Evolution Touch can homogenize a large number of this document, Bertin Technologies disclaims all liability for any resulting different samples in between a few seconds and two minutes.
  • Page 4: Recap Of The Advice For Using The Power Supply

    Heavy load hazard: there is a heavy load sticker on the packaging The means of disconnection of the equipment is the mains of the equipment. The Precellys Evolution Touch should be connector. carried by two operators, preferably. Safety shoes must be worn Disconnecting the power cord from the device cuts off the by the operator during transportation.
  • Page 5: Biological Risks

    Protective glasses Gloves 2.5. Non-ionising radiation Bertin Technologies shall be exempt from any claim for compensation for any damage to property or physical injury resulting from a non-compliance WARNING: with the safety instructions set out in this document. This device is a Class A apparatus. In a residential environment, this device may cause radio interference.
  • Page 6 USER MANUAL PRECELLYS Evolution Touch ® Lid: closes the equipment. 3.1.2 Tube kinetics Locking handle: locks the Precellys® Evolution Touch lid and Given the design and symmetry of the machine, each of the tubes of prevents it from opening. the same type placed in the rack has strictly the same kinetics, which guarantees the same quality of lysis and homogenization in each of the Human Machine Interface: allows the user to program and tubes.
  • Page 7 USER MANUAL PRECELLYS Evolution Touch ® 3.1.4.1 Programming 3.1.4.2 Description of the command interface The PRECELLYS® Evolution Touch is designed to operate at a maximum speed of 10 000 rpm. A PRECELLYS® Evolution Touch operating protocol is made up of several successive cycles.
  • Page 8 USER MANUAL PRECELLYS Evolution Touch ® 3.1.4.3 Setting up protocols on the Precellys® Evolution Touch Icon Values Protocol settings The Precellys® Evolution Touch operates at a maximum speed of10000 rpm. Grinding time 5 to 120 s A homogenization protocol for a Precellys® Evolution Touch is made up of one or more steps between which the lid must not be opened.
  • Page 9 USER MANUAL PRECELLYS Evolution Touch ® 1. Enter a protocol name using the virtual keyboard of the protocol are set, launch the protocol by pressing the start button of the 2. Select icon to be associated with the protocol. instrument (below the screen) or save the modified protocol under another name.
  • Page 10 ® 3.1.4.8 Access to the settings of the Precellys Evolution Touch device This menu is composed of a list of spare parts and specifies the maintenance actions that users must perform on a regular basis in order to ensure the In order to modify the setting of the device, press the setting icon on the continuous good working order of Precellys®...
  • Page 11: Characteristics

    70 dB. Technical characteristics Warranty Supply voltage and frequency 110 V~ / 230 V~ (+/- 10%)- 50 Hz / 60 Hz Bertin Technologies warrants that the equipment is free of defects when it Consumption <1 kVA is shipped. Fuse 5x20 –...
  • Page 12: Standardisation Requirements

    USER MANUAL PRECELLYS Evolution Touch ® Standardisation requirements TRANSPORT AND STORAGE This equipment complies with the requirements of the CE marking except Transport for requirement 6.3.2 b) of IEC 61010-1. However, with respect to this requirement, the equipment complies with the Low Voltage Directive and Avoid violent shocks which could make the device not work properly.
  • Page 13: Installation And Connection

    USER MANUAL PRECELLYS Evolution Touch ® 2. Carefully remove the Precellys® Evolution Touch from its box and place Installation and connection it on a clean, horizontal and stable work surface (weight of the device = 27 kg): the height of the 4 feet can be adjusted in order to make the 1.Remove the foam pad from the tube holder and replace the containment equipment more stable.
  • Page 14: Operatinginstructions

    4. Check that the air intake of the fan and the base of the casing are Only the lysing kits recommended by Bertin Technologies should be used. unobstructed: the air outlet is located under the device. Using any other accessories may reduce the level of protection provided by the device.
  • Page 15 USER MANUAL PRECELLYS Evolution Touch ® Note: When loading tubes, ensure that: The vacuum joint is correctly positioned in its slot on the holder. The blocking plate is centered on the holder and rests on the Figure 2 : 2mL holder and vacuum joint.
  • Page 16: Alarms

    Pressure error 1.Switch off the device. The vacuum circuit is defective. The lysing kits produced by Bertin Technologies have different usage 2.Contact the technical support. limitations and must be used within these limits. The litis are indicated on the www.bertin-instruments.fr website.
  • Page 17: Cleaning And Desinfection

    (vents and cleaning and/or decontamination was performed. (see appendix 1). air inlet). If the return form is missing from the consignment, Bertin Technologies Unplug the equipment before cleaning. after-sales team reserves the right to refuse to perform repairs.
  • Page 18: Wear Parts Replacement

    USER MANUAL PRECELLYS Evolution Touch ® Wear parts replacement 9.3.1 vacuum joint replacement This paragraph specifies the maintenance that users must perform on a The vacuum joint located on the metal holder wears regular basis to ensure that the Precellys® Evolution Touch works properly out during operation.
  • Page 19: Support

    The equipment, its spare parts and its consumables are to be sorted and recycled in accordance with the directive on waste electrical and electronic equipment (WEEE) 2012/19/EU. Bertin Technologies is responsible for processing WEEE for its products sold in France (via RECYLUM). For other countries, the importer is responsible for waste processing.
  • Page 20 USER MANUAL PRECELLYS Evolution Touch ® 5. Place the blocking 11.2.2 Preparing Precellys® Evolution Touch plate tubes together as in the figure 1. Open the lid and Close the lid and place them into place the black foam the special cutout of padding inside.
  • Page 21 USER MANUAL PRECELLYS Evolution Touch ® NOTES NOTES Page Page...
  • Page 22 USER MANUAL PRECELLYS Evolution Touch ® NOTES NOTES Page Page...
  • Page 23 Precellys 24 Minilys Cryolys Precellys Lysing Kits The Original The Individual The Cooling Unit More than 1000 scientific protocols and data on the application center Join the Precellys community now! Bertin Technologies 10 bis, avenue Ampère, 78180 Montigny-le-Bretonneux, France www.bertin-technologies.com...