Replacement
seca 674, 675
1.2 Replacing the load cells
To replace the load cells (G, Fig. 3), first detach the frame from the platform. Remove
the platform mat to do so.
Then undo the four screws (H) that connect the platform with the frame.
You can now take out the frame.
Proceed as described under 1.1 and open the housing for the evaluation electronics
(C, Fig. 1).
Unsolder the connection from the load cell to the NEC2 weight module (F).
Refer to cabling diagram 08-02-06-058.
Unscrew the adjustable feet (I, Fig. 3).
Turn round the frame, undo the 4 screws (J) that secure the load cell to the frame (K)
(Fig. 3).
Carefully pull out the load cell with the load cell cables from the frame tubes.
Proceed in the reverse order for reassembling.
When reassembling refer to cabling diagram 08-02-06-058.
H
H
Figure 2: Top view scale
08.02.2013/MRE
Seite 5 (9)
H
H
30-34-00-859_E1