Overview Of Adjusting A Baby Scale With 10G Graduations [374|376|378] - Seca 378 Service Manual

Hide thumbs Also See for 378:
Table of Contents

Advertisement

Maintenance
2.1.9 Overview of Adjusting a Baby Scale with 10g Graduations
[374|376|378]
Action carried out by the user
The scale is switched off.
Press the tare/hold key and keep it
pressed while simultaneously pressing
the start key.
Release both keys.
Press the tare/hold key for more than 1.5
seconds.
18 seconds have passed and no key has
been pressed.
Press the tare/hold key for more than 1.5
seconds.
Place the test weight on the scale.
Decide whether to decrement or to
increment:
Incrementing:
Press the tare/hold key for more than 1.5
seconds.
Decrementing: No action required as the
scale already is in dec mode.
Decrementing or incrementing:
14.11.12/RLEU/ARI/MRE
Result
No values or characters on the display.
The contents of the calibration counter
flash on the display for 18 seconds.
The contents of the calibration counter
flash on the display for 18 seconds.
The system switches to adjustment
selection mode and "CAL" flashes on the
display.
The scale switches off.
"dec" appears on the display (flashing)
and then the current weight (0.000 - also
flashing).
The measured weight of the test weight
appears on the display (flashing).
Incrementing mode is activated; "inc"
appears on the display.
No output on the display.
The measured weight is decremented or
16/40
seca
30-34-00-812c

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents