Radio Error Symptoms [374|376|378|703|704|284|285|634|635|644|645|656|657|664|665| 674|675|676|677|684|685|954|957|959|963] - Seca 378 Service Manual

Hide thumbs Also See for 378:
Table of Contents

Advertisement

Errors
3.3
Radio Error Symptoms
[374|376|378|703|704|284|285|634|635|644|645|656|657|664|665|
674|675|676|677|684|685|954|957|959|963]
Error description
Device does not
transfer any data
when the SEND
or PRINT key is
pressed
Device beeps
twice when
sending and
printing
The device
displays "Wait" at
the end of the
teach-in phase
14.11.12/RLEU/ARI/MRE
Possible causes
Teach-in phase not
completed out successfully
Receiving device not
switched on
Two devices are in the
teach-in phase at the same
time
23/40
seca
Remedy
Repeat teach-in
phase (see operating
instructions)
Switch on all devices
in the radio network
Switch off all devices
within radio
transmission range
and repeat the teach-
in process
30-34-00-812c