Updating the Radio Module Software

Two cases can be distinguished:
1) The customer has a seca 360° wireless USB adapter [456] and has
carried out set-up for his seca devices and this wireless USB adapter.
In this case you should connect the customer's wireless USB adapter to
your service PC and update the entire wireless group from your PC.
2) The customer does not have a wireless USB adapter.
In this case you can use your own wireless USB adapter. Prior to
updating, set-up must be performed for the customer's devices.
Note!
The customer's wireless group will be overwritten by the set-up
process. After updating, set-up of the customer's devices must be
repeated – without connecting your wireless USB adapter. Teaching in
again is necessary since otherwise the customer would receive error
messages in future informing him that the wireless USB adapter is not
available.
Information regarding the currently active wireless group is displayed in the
main window of the seca wireless updater. In the example shown, wireless
group 2 is active in which set-up has been performed for a wireless USB
adapter, a baby scale and a wireless printer.
The seca wireless updater attempts to establish a wireless connection to the
individual devices. Responding devices are indicated by a green symbol,
devices in the wireless group which have been taught in but which are currently
not available – e.g. because they are switched off – are indicated by a red
symbol. A blue symbol means that the corresponding device is in updating
14.11.12/RLEU/ARI/MRE