Error Symptoms For Head Slider [374|376|378|703|704|284|285|954|957|959|963]; Radio Error Symptoms [374|376|378|703|704|284|285] - Seca 7031321007 Service Manual

Table of Contents

Advertisement

Errors
3.2
Error Symptoms for Head Slider
[374|376|378|703|704|284|285|954|957|959|963]
Error description
Head slider does
not start
Head slider does
not display a
length or an
incorrect length
3.3

Radio Error Symptoms [374|376|378|703|704|284|285]

Error description
Device does not
transfer any data
when the SEND
or PRINT key is
pressed
05.10.12/RLEU/ARI/MRE
Possible causes
Battery not inserted or flat
Voltage supply defective
Key panel defective or not
connected
Insert scale defective /
missing / scale top and
bottom reversed
Head slider not adjusted
correctly
Light barriers soiled
Key unit not engaged
correctly
Possible causes
Teach-in phase not
completed out successfully
20/38
seca
Remedy
Check the batteries
See section 4.1
Check cable
connections, check
cable and key panel
for interruptions,
replace if necessary
Check the insert scale
See section 2.2
Clean the head slider
barriers with a cloth
Check the key unit
Remedy
Repeat teach-in
phase (see operating
instructions)
30-34-00-812b