Display Refresh Time (08. Menu Item) - Datcon DT9000 N Operating Instructions Manual

Process indicator
Table of Contents

Advertisement

DT9000 N
Function
Sequence of operations 1. Enter the menu by pressing a supervisor code.
20180614-V0

7.10. Display refresh time (08. menu item)

The instrument performs cca. 15 measurements in each
seconds. It's too fast to see the changing of the
measurement value.
Here you can define the time periods by which the
instrument displays the new measurement values.
[Default factory setting: 0.5 second]
The way the code should be type in is described in Chapter
7.1. Type the code in. You see on the display:
2. Keep stepping by pressing the ▲ button as long as you
see this menu item:
3. Enter the menu item by pressing the ENTER button.

4. You see this
seconds) The 0.5 numerical value is blinking.
5. Pressing the ▲ button select the time periods by which
the instrument should display the new measurement values.

= once in each 0.5 seconds [Default factory
setting].

= once in each 1 second.

= once in each 1.5 seconds.

= once in each 2 seconds.

= once in each 0.1 seconds.

= once in each 0.3 seconds.

.
on the display. It's meaning: 0.5

.
47