Select connection cable Take note the suitability of the connecting cable.
Select connection cable In particular, make sure that no potential equalisation
Inserting the cable into
the instrument
20180614-V0
We recommend the use of screened twisted pair cable. The
wire cross-section should be 0.25-1.5 mm
currents flow over the cable screen. In case of grounding on
both sides (for suppress the influence of high frequency
interference signals) this can be achived by use of a
capacitor (e.g. ceramic capacitor 1 nF, 1500 V) or separate
potential equalisation. The low frequency potential
equalisation currents are thus supressed, but the protective
effect against high frequency interference signals remains.
1. Prepare the cable for the connection.
Remove approx. 30 mm of cable mantle, strip approx.
8 mm insulation.
2. Remove the instrument front cover (see chapter 4.3.).
3. Loosen compression nut of the cable entry.
4. Insert the cable into the instrument through the cable
entry.