Acknowledge Code And Error Codes - A&D FZ-i WP Series Instruction Manual

Precision balance
Hide thumbs Also See for FZ-i WP Series:
Table of Contents

Advertisement

18.2. Acknowledge Code and Error Codes

When the "AK, Error code (erCd)" parameter of "Serial interface ( 5if )" is set to "1", the balance
outputs <AK> code or an error code for each command as follows:
<AK> (06h) Acknowledge in ASCII code.
 When the balance receives a command to request data and can not process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to request data and can process it, the balance outputs the data.
 When the balance receives a command to control the balance and can not process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to control the balance and can process it, the balance
transmits the acknowledge code.
Among commands to control the balance, the following transmit the acknowledge code both when
the balance receives the command and when the balance has accomplished the command. If the
command can not be processed properly, the balance transmits an error code (EC, Exx). This error
can be released using the CAL command.
CAL command (When performing the calibration using a calibration weight)
ON command, P command
R command, Z command, T command (When setting the display to zero)
R command example
CAL command example (FZ-i / FZ-iWP / FZ-GD series: Calibration using the
internal mass)
Personal computer
CAL command
C
A
L
Time
C R L F
Balance
Zero display
Confirmation of
AK C R L F
command receipt
AK C R L F
Command completion
Zero display
75