Error Codes - A&D FZ-i WP Series Instruction Manual

Precision balance
Hide thumbs Also See for FZ-i WP Series:
Table of Contents

Advertisement

20.2. Error Codes

Display
Error code
pcs
pcs
pcs
EC, E11 Stability error
The balance can not stabilize due to an environmental problem.
Prevent vibration, drafts, temperature changes, static electricity
and magnetic fields.
Refer to "3. PRECAUTIONS" for details on the operating
environment and "7. RESPONSE ADJUSTMENT" about
adapting the balance to the environment.
To return to the weighing mode, press the CAL key.
Out of range error
The value entered is beyond the settable range.
Re-enter the value.
EC, E17 Internal mass error (Only for the FZ-i
The internal mass application mechanism does not function
properly. Perform the weighing operation from the beginning
again.
EC, E20 Calibration weight error
The calibration weight is too heavy.
Confirm that the weighing pan is properly installed. Confirm the
calibration weight value.
Press the CAL key to return to the weighing mode.
EC, E21 Calibration weight error
The calibration weight is too light.
Confirm that the weighing pan is properly installed. Confirm the
calibration weight value.
Press the CAL key to return to the weighing mode.
Overload error
A sample beyond the balance weighing capacity has been
placed on the pan.
Remove the sample from the pan.
Weighing pan Error
The weight value is too light.
Confirm that the weighing pan and pan support are properly
installed. Press the ON:OFF key two times to return to the
weighing mode. If the error still persists, calibrate the balance.
Sample mass error
The balance can not store the sample for the counting mode or
for the percent mode because it is too light.
Use a sample that is heavier
Unit mass error
The sample unit mass for the counting mode is too light. Storing
and using it for counting will cause a counting error.
Add samples to reach the specified number and press the
PRINT key.
Pressing the PRINT key without adding samples will shift the
balance to the counting mode. But, to acquire accurate
weighing, be sure to add samples.
Description
/ FZ-i WP / FZ-GD
80
series)

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents