Correcting The Internal Mass Value: 2 (Only For Fz-I / Fz-I Wp / Fz-Gd Series) - A&D FZ-i WP Series Instruction Manual

Precision balance
Hide thumbs Also See for FZ-i WP Series:
Table of Contents

Advertisement

8.6.

Correcting the Internal Mass Value: 2 (Only for FZ-i / FZ-i WP / FZ-GD series)

 The FZ-i / FZ-iWP / FZ-GD series balance can correct the internal mass value using an external weight.
Method 2: Calibrate the balance using the external weight. The balance performs automatic
calibration using the internal mass, corrects the internal mass value and stores it in
the balance. The corrected mass value is maintained in non-volatile memory even if
the AC adapter is removed.
 The usable calibration weights and the adjustable range are as follows:
Model
FZ-120i
FZ-120iWP FZ-120GD
FZ-200i
FZ-200iWP FZ-200GD
FZ-300i
FZ-300iWP FZ-300GD
FZ-500i
FZ-1200i FZ-1200iWP FZ-1200GD
FZ-2000i FZ-2000iWP FZ-2000GD
FZ-3000i FZ-3000iWP FZ-3000GD
FZ-5000i
The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.
*1 Can be selected with balance software version 3.10 or later.
Operation
1 With the factory setting, the balance can not correct
the internal mass value. Refer to "9. FUNCTION
SWITCH AND INITIALIZATION" and set the internal
mass value correction switch to 1.
2 In the weighing mode, press an d hold the SAMPLE
key to display ba5fnc .
3 Press the SAMPLE key several times until C5 in
displayed.
If C5 in 2 is not displayed, perform step 1.
4 Press the PRINT key to display Cal 0 .
Refer to "8-5 Calibration Using an External
Weight", to perform the calibration.
5 After calibration, end is displayed. The balance displays
?
1ready
when the weight is removed.
Place the accessory breeze break on the balance and
press the PRINT key.
SET
6 1Cal .
is displayed and the balance corrects the
internal mass value automatically
7 After correcting the internal mass value, the balance displays
CCal .
, and performs the calibration automatically using
the corrected internal mass value.
8 The balance displays end and returns to the weighing mode.
9 Confirm that the internal mass value has been corrected
properly by placing the external weight used for the correction
procedure on the weighing pan.
If the internal mass value has not been corrected properly,
repeat the procedure to adjust the correction value.
Usable calibration weight
100 g
50 g
,
200 g
100 g, 50 g
*1
,
300 g, 200 g
100 g, 50 g
,
500 g
400 g, 300 g, 200 g
,
100 g, 50 g
1000 g
500 g
,
2000 g
1000 g, 500 g
,
*1
3000 g, 2000 g
1000 g, 500 g
,
5000 g
4000 g, 3000 g, 2000 g
,
1000 g, 500 g
21
is
38
Adjustable range
-0.050 g to +0.050 g
*1
*1
-0.50 g to +0.50 g
200 g