Acknowledge Code And Error Codes - A&D BM-20 Manual

Analytical balance bm series
Hide thumbs Also See for BM-20:
Table of Contents

Advertisement

Acknowledge Code And Error Codes

16.3.2.
When the "Serial interface function (
<AK> code or error code to each command as follows:
<AK> (06h)
When the balance receives a command to request data and can not process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to request data and can process it, the balance outputs
the data.
When the balance receives a command to control the balance and can not process it, the balance
transmits an error code (EC, Exx).
When the balance receives a command to control the balance and can process it, the balance
transmits the acknowledge code.
<AK> (06h)
Among commands to control the balance, the following transmit the acknowledge code both
when the balance receives the command and when the balance has accomplished the command.
If the command can not be processed properly, the balance transmits an error code (EC, Exx).
This error can be released using the CAL command.
CAL command
ON command
P command
R command
TR command
TST command
When a communication error has occurred due to external noise, or a parity error has occurred
due to transmission error, the balance transmits an error code (EC, Exx). In this case, send the
command again.
Settings Related To RS-232C
16.3.3.
Concerning the RS-232C, the balance has two functions: "Data output (
(
)". Set each function as necessary.
5if
BM series
Acknowledge in ASCII code.
Acknowledge in ASCII code.
(Calibration command using internal mass)
(Display ON command)
(Display ON/OFF command)
(RE-ZERO command)
(Tare command)
(Calibration test command)
is error code number.
xx
)" parameter is set to "
5if
is error code number.
xx
75
", the balance outputs
erCd on
)" and "Serial interface
dout