8.3.

Calibration Using an External Weight

This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until
displayed, and then release the key.
3 The balance displays Cal
 If you want to change the calibration weight (a list
of usable weights is shown on page 31), press
the SAMPLE key and go to step 4.
 If you use the calibration weight value stored in
the balance, go to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key
RE-ZERO key (To increase the value)
MODE key (To decrease the value)
PRINT key
CAL key
Calout
0 .
To switch the display condition
to: "All of weight selection
mode) or "The last two digits
blinking"
(value
adjustment
mode).
To
select
the
calibration
weight or adjust the value.
To store the new weight
value. Even if the AC adapter
is removed, the data is
maintained in non-volatile
memory.
To cancel the operation and
return to Cal
0 .
33
is
e.g. Calibration
weight value
1000.12 g
Press and
hold the key
(for 2 seconds)
Release
the key
Select a
weight
To next page