Calibration Test Using An External Weight - A&D FZ-i WP Series Instruction Manual

Precision balance
Hide thumbs Also See for FZ-i WP Series:
Table of Contents

Advertisement

8.4.

Calibration Test Using an External Weight

This function tests the balance weighing accuracy using an external weight and outputs the result. This
is available only when the "GLP output (info)" parameter is set to "1", "2"or "3". (Calibration test does
not perform calibration.)
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until
displayed, and then release the key.
3 The balance displays CC
 If you want to change the target weight (a list of
usable weights is shown on page 31), press the
SAMPLE key and go to step 4.
 If you use the target weight value stored in the
balance, go to step 5.
4 Specify the target weight value as follows:
SAMPLE key
RE-ZERO key (To increase the value)
MODE key (To decrease the value)
PRINT key
CAL key
CCout
0 .
To
switch
the
display
condition to: "All of the
segments blinking" (target
weight selection mode) or
"The last two digits blinking"
(value adjustment mode).
To select the target weight or
adjust the value.
To store the new weight
value. Even if the AC adapter
is removed, the data is
maintained in non-volatile
memory.
To cancel the operation and
return to CC
0 .