Correcting The Internal Mass Value - A&D GP-61KS Instruction Manual

Gp series precision balance
Table of Contents

Advertisement

7-6 Correcting the Internal Mass Value

The GP series balance can correct the internal mass value within the range shown below. This
function corrects the internal mass value to conform to an external weight. The corrected mass value
is maintained in non-volatile memory even if the AC adapter is removed.
The internal mass value is corrected as follows:
Model
GP-12K
GP-20K
GP-30K/GP-30KS
GP-32K/ GP-32KS
GP-61K/ GP-61KS
GP-40K
GP-60K/ GP-60KS
GP-100K/GP-100KS
GP-102K
Example: Using the GP-20K
Correction reference value
10 kg
20 kg
40 kg
60 kg
100 kg
32
Correction range
-5.0 g to +5.0 g
-50 g to +50 g
-50 g to +50 g

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents