7-3 Calibration Using an External Weight

This function calibrates the balance using an external weight.
Operation
1
Connect the AC adapter and warm up the balance for
at least 30 minutes with nothing on the pan.
2
Press and hold the
displayed, and then release the key.
3
The balance displays
If you want to change the calibration weight (a list of
usable weights is shown on page 28),
press the SAMPLE key and proceed to step 4.
If you use the calibration weight value stored in the
balance, proceed to step 5.
4
Specify the calibration weight value as follows:
SAMPLE key
RE-ZERO key
MODE key
PRINT key
CAL key
CAL
key until
Cal 0 .
To switch the display condition to:
"All
of
the
segments
(calibration weight selection mode) or
"The last two digits blinking" (value
adjustment mode).
(To increase the value)
(To decrease the value)
To select the calibration weight or
adjust the value.
To store the new weight value. Even
if the AC adapter is removed, the
data is maintained in non-volatile
memory.
To cancel the operation and return to
_Cal 0 .
Calout
is
blinking"
30
Select a
weight
e.g. Calibration
weight value
1000.12 g
To next page
Press and
hold the key
Release
the key