WARNING
The procedures below must be carried out by suitably trained and
qualified personnel. Failure to do so may result in injury or an
unsafe product.
Actions to be done by qualified personnel
Check that the bedstripper (linen shelf)
position when maximum foot down tilt is applied
Check that the bed operates correctly using the backup battery
as described in the section Battery test on page 38.
Check operation of the castors, paying special attention to
braking and steering functions
Check that the calf section moves to the Fowler position when the
thigh section is raised
Check that the calf section latches securely in the horizontal
(vascular) position when manually raised.
Check that the bed extension locks securely in all three positions
Examine the power supply cord and mains plug; if damaged,
replace the complete assembly; do not use a rewireable plug
Examine all accessible flexible cables for damage and
deterioration
Check all accessible nuts, bolts and other fasteners are present
and correctly tightened
Check any accessories fitted to the bed, paying particular
attention to fasteners and moving parts