Fault indications

Indication
ACP mattress platform height and
head down tilt indicators flashing
ACP mattress platform height and
foot down tilt indicators flashing
ACP backrest indicator flashing
ACP thigh section indicator flashing
Mattress platform height, head down
tilt, backrest and thigh section
indicators flashing

Product lifetime

40
The bed's control software indicates problems in the electrical
system by means of flashing indicators on the Attendant
Control Panel (ACP). If you experience any of the indications
below, contact ArjoHuntleigh or an approved service agent.
The lifetime of this equipment is typically ten (10) years.
"Lifetime" is defined as the period during which the product will
maintain the specified performance and safety, provided it has
been maintained and operated in conditions of normal use in
accordance with the requirements in these instructions.
Possible cause
Height actuator fault
(foot end)
Height actuator fault
(head end)
Backrest actuator fault
Thigh section actuator
fault
Control unit fault