Contra-Indications - Arjohuntleigh ENTERPRISE E8000X Instructions For Use Manual

Table of Contents

Advertisement

Contra-
indications
WARNING
If the combined weight of the mattress and accessories exceeds
65kg, the maximum patient weight must be reduced accordingly.
746-585-UK-5
The bed is not suitable for use in the following situations:
• A domestic area, i.e. home healthcare (*Application
Environment 4).
• Outpatient care (*Application Environment 5).
* Application Environments are defined in IEC 60601-2-52.
The bed is not suitable for patients under 40kg in weight.
The maximum recommended patient weight is 185kg.
The safe working load (SWL) of the bed is 250kg.
The safe working load is calculated as follows (in accordance
with IEC 60601-2-52):
Maximum patient weight........................................... 185kg
Mattress .................................................................... 20kg
Accessories (including attached loads) ...................... 45kg
TOTAL
.................................................................. 250kg
The recommended patient height is between 146cm and
190cm.
At the discretion of the caregiver, patients taller than 190cm
may be accommodated by extending the bed - refer to Bed
length adjustment on page 17. Ensure that the patient's
height does not exceed the In-bed length shown on page 43.
5