Summary of Contents for Arjohuntleigh alpha active 4
Page 1
Alpha Active 4 Instructions for Use ...with people in mind 648936EN_04 • 10/2017...
Page 3
• The cover of this product is vapour permeable but not air permeable and may present a suffocation risk. • Only the pump and mattress combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and mattress combinations are used.
Page 4
Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.The content of this publication may not be copied either whole or in part without the consent of ArjoHuntleigh.
If you have any difficulties in setting-up or using the system, contact your local ArjoHuntleigh Alpha Active 4 sales office, listed at the end of this manual.
Alpha Active 4 Mattress Replacement mattress replacement comprises the Alpha Active 4 following components: The standard cover comprises of a 2-way stretch PU Detachable Cover (polyurethane) coated knitted fabric zipped to a durable nylon base. The zips are protected by flaps to prevent ingress of contaminants and allow easy removal of the cover for cleaning.
Page 9
A full technical description of the system Alpha Active 4 can be found in the Service Manual, part number SER0018, available from your ArjoHuntleigh sales office.
Clinical Applications Indications The system is indicated for the Alpha Active 4 prevention and/or management of all categories pressure ulcer, when combined with an individualised, comprehensive pressure ulcer protocol: for example, repositioning, nutritional support, skin care. Selection should be based upon a holistic assessment of the patient’s individual care needs.
Preparing the system Remove the system from the packaging. You should have the following items: for use • pump including mains power cord Alpha Active 4 and hanging brackets. • mattress replacement with integral Alpha Active 4 tubeset. • Cover.
Page 12
WARNING Make sure that the mains power cable is positioned to avoid causing a hazard and is clear of moving bed mechanisms or other possible entrapment areas. The pump should be installed as follows: Installing the Pump 1. Position the pump, feet down, on any convenient horizontal surface or alternatively suspend from the bed foot rail by means of the integral hanging brackets.
4. Controls, Alarms and Indicators Controls Cycle Time Alarm Mute Mode Select Auto-firm Mode Patient Weight Selector Alarms Pressing the button will activate the pump.The On/Off Button On/Off pump will perform a self test during which all LEDs will illuminate, and audible beeps will sound. During an alarm condition the sound of the alarm can be Alarm Mute muted by pressing this button, and the audible alarm will...
Alarms and Indicators indicator is illuminated whenever the Low Pressure Indicator Low Pressure pump detects low pressure within the mattress replacement. An audible alarm will sound, but the audible alarm can be muted by the button. mute The indicator will extinguish once normal pressure is reached.
Do not place the patient on the mattress until it is fully inflated. Quick Start Before using the mattress replacement Alpha Active 4 make sure it has been installed correctly in accordance with “Installing the Mattress” on page 5, and ensure that the CPR unit on the mattress is clicked into the closed position.
To Disconnect the buttons on the top and bottom of the tubeset connector Tubeset pull the tubeset connector away from the pump. To deflate the mattress Refer to “To Deflate and Store the Alpha Active 4 Mattress” on page 11.
Transport mode is non-therapeutic offering support only for up to 12 hours. It is recommended that when in transport mode the patient is frequently monitored. To Deflate and Store the Alpha Active 4 Mattress 1. Disconnect the tubeset from the pump. To deflate the mattress: 2.
CPR Control IMPORTANT IN THE EVENT OF CARDIAC ARREST In the event of a patient suffering cardiac arrest and CPR needing to be administered: To activate CPR Located at the head end of the mattress replacement (on the same side as the tubeset) is a red strap labelled CPR. In the event of a cardiac arrest pull this from the mattress to deflate.
Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist. system should be routinely decontaminated between patients and Alpha Active 4 at regular intervals while in use; as is good practice for all reusable medical devices. WARNING Remove the electrical supply to the pump by disconnecting the mains power cord from the mains power supply before cleaning.
Page 20
For information for the mattress top cover, including Thermal Disinfection laundering guidelines, refer to “Cover Specification” on page 20.
The serial number for the pump is on the label on the Pump back of the pump case. The mattress serial label can be found just inside the Mattress base cover above the tubeset. Refer to “Alpha Active 4 Mattress Replacement” on page 2.
8. Troubleshooting and Alarm Conditions The following table provides a troubleshooting and alarm condition guide for the system in the event of malfunction. These alarms do not cause any Alpha Active 4 delay or interruption in therapy. Indicator Possible Cause...
9. Technical Description PUMP Model: Alpha Active 4 Part Numbers: 648311UK UK (Note: pre 2015 this part number was 648311) 648319AU AUSTRALIA Supply Voltage: 230V Supply Frequency: 50Hz Power Input: 0.1A Size: (L)280mm x (W)205mm x (H)112mm Weight: 2.5kg Case Material:...
PUMP SYMBOLS The operator must read this Power document (Instructions for Power Disconnects Use) before use. O (Off) I (On) Connects to the from the mains Note: This symbol is blue mains supply supply on the product label. With respect to electric shock, fire and mechanical hazards only in accordance with CAN/ Double...
ALPHA ACTIVE 4 MR 90 Polyurethane PVC Nylon Coated ALPHA ACTIVE 4 MR 85 Polyurethane PVC Nylon Coated ALPHA ACTIVE 4 MR 85 (NON-PVC) Polyurethane Nylon PU Coated ALPHA ACTIVE 4 MR 80 (NON-PVC) Polyurethane Nylon PU Coated ALPHA ACTIVE 4 MR 90 (NON-PVC)
Page 26
Chlorine concentrations may vary from 250ppm to 10,000ppm depending on local policy and contamination status. If an alternative disinfectant is selected from the wide variety available, ArjoHuntleigh recommend that suitability for use is confirmed with the chemical supplier prior to use.
Page 27
Guidance and manufacturer’s declaration - electromagnetic emissions The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic environment - guidance RF emissions Group 1...
Page 28
Guidance and manufacturer’s declaration - electromagnetic immunity The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. Immunity IEC 60601 Test Compliance Electromagnetic environment - guidance Test...
Page 29
Guidance and manufacturer’s declaration - electromagnetic immunity The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. Immunity Test IEC 60601 Test Compliance Electromagnetic environment - Level...
Page 31
FRANCE POLSKA AUSTRALIA FRANCE POLSKA ArjoHuntleigh Pty Ltd ArjoHuntleigh SAS ArjoHuntleigh Polska Sp. z o.o. ArjoHuntleigh Pty Ltd ArjoHuntleigh SAS ArjoHuntleigh Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 78, Forsyth Street 2 Avenue Alcide de Gasperi ul.
Page 32
Getinge Group is a leading global provider of products and systems that contribute to quality the three brands of ArjoHuntleigh, Getinge and Maquet Getinge provides solutions for Maquet specializes in solutions, therapies and products for surgical interventions, interventional ArjoHuntleigh focuses on patient handling and hygiene, disinfection, DVT prevention,...