Hach SC4500 User Manual page 37

Profinet and ethernet/ip modules
Hide thumbs Also See for SC4500:
Table of Contents

Advertisement

10. On the Controller organizer tab, select Tasks > Main Task > MainProgram > MainRoutine.
For each Tag add a COP module with the data that follow (refer to the table bellow):
• Source—Add the controller name followed with the I.Data based on the bytes size.
• Dest—Add the parameter Tag. Initial element to be overwritten by the source.
• Lenght—Number of destination elements to copy.
Tags
1 - LDO_Error
2 - LDO_Status
3 - LDO_Measurement_mgL
4 - LDO_Heartbeat
5 - LDO_Temperature
11. Select the option Go Online.
12. Select Download.
13. Select Download.
Type
Integer (INT)
Integer
Float (Real)
Integer
Float
Size
I.Data
2 bytes
0
2 bytes
2
4 bytes
4
2 bytes
8
4 bytes
10