Process Error Messages (Fieldtool); Process Errors Without Message - Endress+Hauser Dosimag Operating Instructions Manual

Electromagnetic flow measuring system for filling applications
Table of Contents

Advertisement

Troubleshooting
26
No.
Type/error message
321
SF: TOL. COIL. CURR
System errors - notice messages
359
SF: PULSE RANGE
502
SN: UP-/DOWNLOAD
ACT.
692
SN: SIM. MEASURAND
8.4

Process error messages (FieldTool)

Type No./error message
PF = Process error Fault message
PN = Process error Notice message
(see also Chap. 8.6, Response of outputs to errors)
421
PF: FLOW RANGE
8.5

Process errors without message

Symptoms
Measured-value reading shown on
display, even though the fluid is at a
standstill and the measuring tube is
full.
The fault cannot be rectified or
some other fault not described
above has occurred.
In these instances, please contact
your Endress+Hauser
service organization.
Cause
Sensor:
Coil current is out of tolerance.
Pulse output:
The pulse output frequency is
outside the permitted range.
• PASSIVE-POSITIVE
PASSIVE-NEGATIVE
f > 1 : (2 ⋅ pulse width)
• AUTO. PULSEWIDTH
f > 12.5 kHz
Device data are being uploaded or
downloaded by means of a control
unit. It is not possible to execute
any other functions.
Simulation of measurand active
(volume flow)
Cause
The volume flow is too high.
Rectification
1. Check the fluid for presence of gas bubbles.
2. Activate the "LOW FLOW CUTOFF" function, i.e. enter or increase the
value for the switching point.
The following options are available for tackling problems of this nature:
Request the services of an Endress+Hauser service technician.
If you contact our service organization to have a service technician sent out,
please be ready with the following information:
– Brief description of the fault
– Nameplate specifications (→ Page 6): Order code and serial number
Returning devices to Endress+Hauser
The procedures on Page 4 must be carried out before you return a flowmeter
requiring repair or calibration to Endress+Hauser. Always enclose a duly
completed "Declaration of contamination" form with the flowmeter. You will
find a preprinted "Declaration of contamination" form at the back of these
Operating Instructions.
Replace transmitter electronics
Components in the electronics defective → order spare parts → Page 28
Remedy / spare part
– Measure the coil resistance.
– Replace sensor or transmitter if necessary.
→ Page 29
1. Increase the setting for pulse value.
and/or
2. Reduce flow
3. Reduce pulse width.
Wait under the process is completed. The
device is restarted automatically.
Switch off simulation.
Remedy / spare part
Reduce flow
and/or
Select a bigger DN.
Endress+Hauser