Malfunctions
9
Malfunctions
9.1
Rectifying malfunctions
Problem
Pump does not start; none of the built-in
LEDs on the TC 400 lights up
Pump does not start; green LED on the TC
400 is flashing
Pump does not attain the final rotational
speed within the specified run-up time
Pump does not attain the ultimate pres-
sure
Unusual noises during operation
Red LED on the TC 400 is on
34
If malfunctions on the pump occur, you will find possible causes and instructions for re-
pair in the following table.
Possible causes
● Electrical supply interrupted
● Operating voltage incorrect
● No operating voltage applied
● TC 400 defective
● At operation without control panel:
Pins 1-3, 1-4 or 1-14 are not connected on
the "remote" connection
● At operation via RS-485: Bridge at Pin 1-
14 prevents control commands
● At operation via RS-485 (without mating
plug): Parameters of the electronic drive
unit not set
● Voltage drop in the cable too high
● Fore-vacuum pressure too high
● Leak
● Gas throughput too high
● Rotor runs hard, bearing defective
● Setpoint for run-up time to low
● Thermal overload:
– Lack of air ventilation
– Water flow too low
– Fore-vacuum pressure too high
– Ambient temperature too high
● Pump is dirty
● Vacuum chamber, pipes or pump are
leaky
● Bearing damage
● Rotor damage
● Splinter shield or protective screen are
loose
● Collective fault
1)
If no Pfeiffer Vacuum display and control unit is available, please contact the Pfeiffer Vacuum Ser-
vice.
Remedy
Check plug contacts on the power supply
Check supply lines on the power supply
Check the output voltage (24 / 48 V DC) on the
connection "DC out" of the power supply
Check the plug contacts on the TC
Apply correct operating voltage
Observe the rating plate
Apply operating voltage
Exchange the TC 400
Contact the Pfeiffer Vacuum Service
Connect the pins 1-3, 1-4 or 1-14 on the "re-
mote" connection
Install the mating plug (scope of delivery) to the
"remote" connection.
Remove the bridge betwenn pins 1 and 14 from
the "remote" connection
Remove the mating plug from the "remote" con-
nection.
Set parameters [P: 010] and [P: 023] "ON" via
interface RS-485.
Please refer to operating instructions of the
electronic drive unit.
Use a suitable cable
Ensure function and suitability of the backing
pump
Perform leak detection
Check sealings and flange fastenings
Eliminate leaks
Reduce process gas supply
Check the bearing for noises
Contact the Pfeiffer Vacuum Service
Extend the run-up time via DCU, HPU or PC
Reduce thermal loads
– Ensure adequate cooling
– Ensure the cooling water flow
– Lower the fore-vacuum pressure
– Adjust ambient conditions
Bake out the pump
Cleaning in case of heavy contamination
– Contact the Pfeiffer Vacuum Service
Leak detection starting from the vacuum cham-
ber
Eliminate leaks
Contact the Pfeiffer Vacuum Service
Contact the Pfeiffer Vacuum Service
Correct the seat of the splinter shield or the pro-
tective screen
Observe the installation notes
Reset by switching the mains OFF/ON
Reset via pin 13 on the "REMOTE" connection
Differentiated malfunction display is possible via
1)
RS-485
Contact the Pfeiffer Vacuum Service