Peripheral Checks
4-3-6-1
DVR Test (if equipped)
Required: Blank [DVD+RW] media.
1.) Make sure the tab menu of "Video" is found on the right-most touch panel tab. If not found, go to
"Utility" -> "Application" -> "Setting" and check "Show video Tab".
2.) Insert DVD+RW. Go to DVR Touch Panel menu.
3.) Press "REC-Standby" to go stand-by.
4.) Start Recording. Perform image scanning including Doppler sound.
5.) Stop Recording.
6.) Exit from stand-by.
7.) Play.
8.) Verify that Recorded image and sound can be played back.
4-3-6-2
Black and White Printer Test
1.) Print an image by pressing P button associated with BW Printer. (P key should be configured
correctly.)
2.) Verify that image can be printed.
3.) Verify that printed image has no distortion as compared the image on screen.
4-3-6-3
Color Printer Test
1.) Print an image by pressing P button associated with Color Printer. (P key should be configured
correctly.)
2.) Verify that image can be printed.
3.) Verify that printed image has no distortion as compared the image on screen.
4-3-6-4
CD/DVD Read/WriteTest
1.) Insert Blank CD/DVD Media into Optical Drive.
2.) Go to Utility à System à Backup Restore. Select "CD/DVD" in "Media" list.
3.) Select "User Define Configuration checkbox" in Backup and Press "Backup".
4.) Start backup into Media by pressing "OK" Button.
5.) Verify that "Finished OK" message appears.
4-3-6-5
ECG Test
1.) Connect a any probe. Select cardiac application like "Adult". And start scan.
2.) Verify that ECG timeline gets displayed on image area.
3.) Enter CSD.
4.) Select DIAGNOSTICS - Service Diag - ECG Test.
5.) Verify diagnostics exhibit "PASS".
4-3-6-6
V-NAV Test
1.) Enter CSD.
2.) Select DIAGNOSTICS - Service Diag - Drive Bay Test - SYSTEM.
3.) Verify diagnostics exhibit "PASS".