Connecting the Utilities

Connecting the Power

Plug the SYSTEM POWER cord on the back of the DXF to an electrical outlet rated in accordance with
the power input information on the back of the instrument.
CAUTION: Whenever plugging or unplugging power cords, handle them by the plugs, not by the
cords.
MISE EN GARDE: Chaque fois que vous branchez ou débranchez les cordons d'alimentation,
tenez-les par les fiches et non par les cordons.
CAUTION: An independent heavy GROUND wire must be provided through the power hook up.
Improper grounding may cause severe damage for which the supplier will not accept responsibility.
All power strips must be fully grounded and carry the ground through to the sockets into which the
computer is plugged.
MISE EN GARDE: Un fil de masse lourd et indépendant doit être fourni via le raccordement élec-
trique. Une mise à la terre incorrecte peut causer de graves dommages pour lesquels le fournisseur
décline toute responsabilité. Toutes les barrettes d'alimentation doivent être entièrement mises à la
terre et acheminer la terre jusqu'aux prises dans lesquelles l'ordinateur est branché.

Connecting the LN2 Lines

1
Obtain an LN2 tank that can provide a maximum of 22 psi per the Thermal Diffusivity DXF 200 Site
Preparation Guide.
2
Connect the LN2 LIQUID OUTLET on the tank to the LN2 INLET fitting on the right side of the EM.
See
Figure
3.
3
Connect the CRYOGENIC VALVE cable to the LN2 OUTLET on the top of the DXF (see
Figure 10 LN2 tank connections.
DXF-200 Getting Started Guide
CRYOGENIC VALVE
LN2 LIQUID OUTLET
Figure
9).