Chapter 4:
The primary maintenance procedures described in this chapter are the customer's responsibility. Any fur-
ther maintenance should be performed by a representative of TA Instruments or other qualified service per-
sonnel. Consult the Help documentation installed with the instrument control software for further
information.
DANGER: Because of the high voltages in this instrument, untrained personnel must not
attempt to test or repair any electrical circuits.
DANGER: À cause de la présence de tensions élevées dans cet instrument, le personnel non
formé ne doit pas essayer de tester ou de réparer les circuits électriques.
CAUTION: Before using any cleaning or decontamination method except those recommended by
TA Instruments, check with TA Instruments that the proposed method will not damage the instru-
ment.
MISE EN GARDE: Avant d'utiliser une méthode de nettoyage ou de décontamination autre que
celle recommandée par TA Instruments, vérifiez auprès de TA Instruments que la méthode proposée
n'endommagera pas l'instrument.

General Cleaning Practices

Use protective eye glasses
Use finger coats or gloves
Always hold optical components (mirrors, lenses, windows, etc.) from the edges; never touch the cen-
ter after cleaning
If strongly burned films remain on the components after the cleaning process, use weak acid (like vin-
egar) to try to dissolve the residual layers
Always use distilled water and acetone at the end of cleaning process
Use a thin layer of vacuum grease to seal all O-ring parts of the EM module whenever it is necessary to
reassemble the block containing O-rings
DXF-200 Getting Started Guide
Maintaining the DXF/EM