DXF Top Panel
The DXF top panel has three ports; the table below provides a description of the function of each port.
Refer to Figure 7 for an illustration of rear connections.
Port
TEMPERATURE SENSORS
HEATER POWER
CRYOGENIC VALVE CONNECTION
2
2
1
Figure 9
Environmental Module top panel connections.

Connecting the Environmental Module to the Pulse Source

To connect the DXF to the EM, follow the instructions below. Refer to
connection ports, or the diagram located in the Appendix.
1
Connect the SOLID STATE DETECTOR cable from the back of the head plate on the EM to the J102
port on the back of the DXF.
2
Connect the TEMPERATURE SENSORS cable from the side of the EM to the
of the DXF.
3
Connect the HEATER POWER cable from the back of the EM to the
4
Connect the CRYGENIC VALVE CONNECTION from the LN2 OUTLET on the tank to the
port on top of the DXF.
DXF-200 Getting Started Guide
Function
Connect Furnace and Sample RTD wires to Controls
Connect Heater Wires to Controls
Connect the Cryogenic Cooling Valve to the Controls
3
SOLID STATE
DETECTOR cable to the
DXF
PURGE GAS tubing
connected to the PURGE
GAS INLET
TEMPERATURE
SENSORS (Sample and
Control)
HEATER POWER
CRYOGENIC VALVE
CONNECTION
Figure 8
and
Figure 9
above for
port on the top
3
port on top of the DXF.
2
1