Section 7
Initial Operation
System Cleaning
Procedure (cont.)
7-4
Nanopure TOC - UV
Clean your Nanopure as follows:
1. (If you have installed an Accudispense volumetric remote dispenser,
place it in manual mode.) From the (Idle) display, press the UP or
DOWN arrow until display reads, "Clean Unit?"
2. Press ENTER.
3. Unit display will ask if you have "Disconnected Inlet Water (NO)."
4. Disconnect feed water supply at the quick-disconnect inlet valve on left
side of unit.
5. Press the UP or DOWN ARROW to select (YES) and then press
ENTER. Display will now read, "Injected Cleaner (NO)."
6. Open front (left side door) to expose cartridge pack. Remove solution
syringe from wrapper.
7. Remove luer cap on system injection port, located on the right side of
the cartridge pack manifold by turning counterclockwise.
Figure 7-2. Injecting Cleaning Solution
Warning Avoid splashing cleaning solution on clothing or skin.
Ensure all piping connections are tight to avoid leakage.
Ensure adequate ventilation.
Carefully follow manufacturer's safety instructions and material safety
data sheets.
This device to be used with water feeds only.
Cleaning agents must be used in compliance with instructions in this
manual. Failure to comply could result in explosion and personal
injury.