Error Codes - IKA KS 130 control Operating Instructions Manual

Hide thumbs Also See for KS 130 control:
Table of Contents

Advertisement

Available languages

Available languages

AS 130.4 Reactant glass adapter
g
2500
2000
1500
KS 130
1000
500
0
0
80
160
240
320
400
AS 260.1 Universal attachment
g
8000
7000
6000
5000
4000
HS 260
3000
2000
1000
0
200
250
0
50
100
150
AS 260.5 Separating funnel attachment
g
8000
7000
6000
5000
4000
HS 260
3000
2000
1000
0
0
50
100
150
200
250
rpm
1/min
480
560
640
720
800
KS 260
rpm
1/min
300
350
400
450
500
KS 260
rpm
1/min
300
350
400
500
450

Error codes

If an error code is encountered, an audio warning signal is generated and the error
code appears on the display field. First, please try to continue operating by turning
the device off and back on again. If it is still not possible to eliminate an error after
an extended wait, please contact our service department. If you do so, you should
always tell us what error code has been encountered. This simplifies the process
of tracking down the error and makes it possible to form a preliminary opinion.
Error code
Error
Er 2
No communication between PC and
control device in remote operation with
active watchdog function in Mode 1.
Er 3
Internal device temperature is too
high.
Er 4
Motor locked or over loaded, problem
with reading data/signal from light
barrier.
Er 9
Error while reading stored values.
Er 41
Triac defective
Er 42
Safety relay defective
WD (
)
No communication between PC and
control device in remote operation with
active watchdog function in Mode 2.
Cause of the error
- Interface not connected.
- PC is not sending any data within the
amount of time set by the watchdog.
- Permissible ambient temperature
exceeded.
- Shaking table is being hindered in its
stroke motion
- Internal device error
-PCB Logic.
-Internal device error.
-Internal device error.
- Interface not connected
- PC is not sending any data within the
amount of time set by the watchdog.