AS 260.2 Holding bracket adapter
With additional holding brackets (AS2.1, AS2.2, AS2.3, AS2.4 and AS2.5) the
holding bracket adapter is suitable for working with round flasks, measuring flasks
and Erlenmeyer flasks.
Recommended / maximum fitting
AS 2.1 Holding bracket
AS 2.2 Holding bracket
AS 2.3 Holding bracket
AS 2.4 Holding bracket
AS 2.5 Holding bracket
ATTENTION! The holding bracket adapter is secured in place by four lateral
clamping screws on the agitation table.
AS 260.3 Dish attachment
For gentle shaking at low speed, for example for Petri dishes or culture flasks.
The dish attachments is equipped with an anti-slip sheet that prevents the dish
from moving around during the agitation process.
ATTENTION! The dish attachment is secured in place by four lateral clamping
screws on the agitation table.
AS 260.5 Separating funnel attachment
For intensive shaking of separating funnels.
Max. fitting:
6 x 50 ml Separating funnel
5 x 100 ml Separating funnel
3 x 250 ml Separating funnel
3 x 500 ml Separating funnel
ATTENTION! The separating funnel attachment is secured in place by four
lateral clamping screws on the agitation table.
26
12 / 12
Dimension (WxHxD) [mm]
8 / 9
330 x 24 x 334
5 / 8
4 / 4
2 / 4
Dimension (WxHxD) [mm]
410 x 33 x 370
Dimension (WxHxD) [mm]
334 x 145 x 425

Supported weight (load)

In order to ensure safe and secure operation, the shaker must only be operated
within the range of usable weight and speed (see diagram - hatched area).
Please make certain that the holding surface on which the shaker is placed
is clean and level. Otherwise the working range shown in the diagram (the
hatched area) cannot be used.
If the shaker is screwed together with an adjacent piece of equipment (i.e. if it
is integrated into a system) it should be noted during operation that dynamic
forces may arise as a result of unfavorable loading circumstances and the posi-
tion of the center of gravity. This may cause the adjacent equipment to vibrate
or may cause the table to vibrate uncontrollably. If you notice that the device
is running unevenly, the speed must in any case be reduced until the running
irregularities no longer occur.
Make certain that individual agitation vessels are placed in the middle of the
shaking table and multiple shaking vessels are place evenly on the shaking
table, and that all vessels are securely fastened in place.
AS 130.1 Universal attachment
g
2000
1750
1500
1250
KS 130
1000
750
500
250
0
0
80
160
240
320
400
rpm
1/min
480
560
640
720
800